Technical specifications
|

|
|
Infusion
|
|
|
Overall Accuracy*
|
+/- 2 % (Measured time > I hr, min 0.2 ml)
|
|
Pumping Mechanism
|
Lead screw mechanism
|
|
Flow rate ranges
|
1.-200 ml/hr for 50/60 ml syringe
|
|
|
1.-.120 ml/hr for 30 ml syringe
|
|
|
1. -100 ml/hr for 20 ml syringe
|
|
|
0.1 -060 ml/hr for 10 ml syringe
|
|
Infusion volume limits
|
60 ml, Programmable every 0.1 ml
|
|
Bolus rate
|
Up to 1000 ml/hr
|
|
Occlusion pressure limit
|
Programmable three level
|
|
|
Low 0.4 +/- 0.08 Kg/cm² (bar) Medium 0.9 +/- 0.20 Kg/cm² (bar) High 1.2 +/- 0.25 Kg/cm² (bar)
|
|
KVO rate
|
1 % of selected infusion rate but min.0.1 ml/hr.
|
|
Syringe type
|
BD
|
10/20/50/60 ml
|
|
|
Dispovan
|
10/20/50 ml
|
|
|
Terumo
|
10/30/50/60 ml
|
|
|
Pharma
|
10/50 ml
|
|
|
Prefuser
|
50 ml
|
|
Indicators
|
Run & Mains.
|
|
Alarms
|
No mains, Very low battery, Low battery, Occlusion No syringe, Idle, Near End, End of infusion, KVO
|
|
Power
|
|
|
External power supply
|
AC Input range 170 -270 VAC, Frequency 50 -60 Hz
Output: 12 VDC -1 Amp
|
|
Power Consumption
|
12 Watts max.
|
|
Build in battery
|
Ni -MH battery for increased durability.
|
|
Battery life
|
Continuous operation for about 20 hrs. at an operating rate of
5 ml/hr, under standard conditions.
|
|
Classification
|
Class 11, IPXI (Drip Proof)
|
|
Operating Conditions
|
Ambient Temperature
|
05 to 40°C
|
|
|
Relative Humidity
|
20 to 90 % (no condensation)
|
|
|
Atmospheric pressure
|
70 to 106 kPa
|
|
Housing & construction
|
|
|
Compact and Robust F.R.P
|
|
|
Weight
|
Approx. 2.2 Kgs.
|
|
Dimensions
|
Approx. 265 (L), 110 (H), 150 (W)
|
| Mounting |
On a pole or bed-frame |
|
- *As per IEC standard 60601-2-24
Specifications and appearance are subject to changes without prior notice.
|
|
Code
|
Item
|
|
|
2575
|
Syringe Infusion Pump
|
|
|
Request
Quote
|
|
|
| Drop Infusion Pump (TIASHI - DP) |
|
|
Drop Infusion Pump
Technical specifications
|

|
|
Infusion
|
|
|
Overall Accuracy*
|
+/- 1 %
|
|
Setting Modes
|
Flow rate only, flow rate + volume target, time + volume target
|
|
Flow rate ranges
|
0 -300 drops/min or 0 -300 ml/hr.
|
|
Infusion volume limits
|
0 -60000 drops or 0 -10000 ml
|
|
Time range
|
0 -99:00 hrs
|
|
KVO rate
|
1 to 5 % of selected infusion rate but min. 1 drop/ min or 1 ml/hr
|
|
Mechanical
|
|
|
Pumping Mechanism
|
Linear Peristaltic
|
|
Drop Sensor
|
External
|
|
Air Bubble Detector**
|
Ultrasonic Adjustable 20- 150µ1
|
|
Set type
|
15, 20, 40, 60 drops/cc.
Facility to add four customized sets as per user's choice.
|
|
Indicators
|
Run & Mains.
|
|
Alarms
|
No mains, Very low battery, Low battery, Sensor error, Low flow, High flow, Door open, Near End, End of infusion, KVO, Air in line**, System error.
|
|
Special Features
|
|
|
Free Flow Protection
|
A unique lever mechanism protects from free flow in case of accidental door opening.
|
|
Digital Processing of Drop Sensor Signal
|
Digital processing of drop sensor signals allows you to the change the infusion parameters without the need for infusion interruption
|
|
Adjustable air in line alarm threshold **
|
The size of air bubble to be detected can be set in the range of 20 to 150µ1
|
|
Flow Rate & Volume Unit selection
|
Option to select Flow Rate Unit in terms of drop/mill or ml/hr & Volume Unit in terms of drops or m1.
|
|
Power
|
|
|
External power supply
|
AC Input range170 -270 VAC, Frequency 50 -60 Hz
12 Volts, l Amp DC output Supply
|
|
Power Consumption
|
12 Watts max.
|
|
Build in battery
|
8.4 V, 2.1 AHr Ni -MH battery for increased durability.
|
|
Battery life
|
Continuous operation for about 4.5 hrs. at an operating rate of 20 drops/mill or 20 ml/hr with assign I. V. Set under standard conditions.
|
|
Classification
|
Class II, IPX1 (Drip Proof)
|
|
Operating Conditions
|
Ambient Temperature
|
05 to 40° C
|
|
|
Relative Humidity
|
20 to 90 % (no condensation)
|
|
|
Atmospheric pressure
|
70 to 106 kPa
|
|
Housing & construction
|
|
|
Compact and Robust A.B.S Plastic
|
|
|
Weight
|
Approx. 2.4 Kgs.
|
|
Dimensions
|
Approx. 250 (L), 245 (H), 90 (W)
|
|
Mounting
|
On a pole or bed-frame
|
|
|
|
* As per IEC standard (i0601-2-24
** Optional feature
Specifications and appearance are subject to changes without prior notice.
|
| Code |
Item |
|
2576 |
Drop Infusion Pump |
|
|
|
|
|
Request